Ability of the third-generation FloTrac/Vigileo software to track changes in cardiac output in cardiac surgery patients: a polar plot approach.

Abstract : OBJECTIVE: To evaluate the ability of the third-generation (3.01) of FloTrac/Vigileo monitor (Edwards Lifesciences, Irvine, CA) to follow variations in cardiac output (∆CO) using the new polar plot approach. DESIGN: Prospective interventional study. SETTING: Single hospital university study. PARTICIPANTS: Twenty-five patients referred for cardiac surgery. INTERVENTIONS: CO was measured simultaneously by 3 to 5 bolus thermodilution (COtd measurements), using a pulmonary artery catheter and by arterial pulse contour analysis, using the FloTrac/Vigileo (COvi). Data were collected at eight time points: before incision, after sternotomy, before and after protamine sulfate infusion, at the start of sternal closure, at the end of surgery, on arrival to intensive care unit, and after a standardized volume expansion with 500 mL of hetastarch 6%. MEASUREMENTS AND MAIN RESULTS: One-hundred thirty-five pairs of CO data were collected; the mean bias of all CO measurements corrected for repeated measures was 0.2 L/min with limits of agreements of -3.3 L/min and +2.9 L/min. The percentage error was 66.5%. The polar plot analysis included 71 significant ∆CO and showed a mean polar angle of -3.4 degrees with 95% polar percentage error equivalent limits of -61 to 55; 69% of analysed data points fell within the 30-degree limits and provided a correct polar concordance rate. CONCLUSIONS: Third-generation FloTrac/Vigileo software still lacks the accuracy to reliably detect changes in cardiac output (∆CO) in cardiac surgery. Improvements to FloTrac/Vigileo CO algorithm and software still are needed in this particular setting.
Type de document :
Article dans une revue
Lyon Chir, 2013, 27 (6), pp.1122-7. 〈10.1053/j.jvca.2013.03.008〉
Liste complète des métadonnées

https://hal-hcl.archives-ouvertes.fr/hal-00979685
Contributeur : Frédéric Riondet <>
Soumis le : mercredi 16 avril 2014 - 15:37:40
Dernière modification le : lundi 12 mars 2018 - 12:26:01

Identifiants

Collections

Citation

Olivier Desebbe, Roland Henaine, Geoffray Keller, Catherine Koffel, Hans Garcia, et al.. Ability of the third-generation FloTrac/Vigileo software to track changes in cardiac output in cardiac surgery patients: a polar plot approach.. Lyon Chir, 2013, 27 (6), pp.1122-7. 〈10.1053/j.jvca.2013.03.008〉. 〈hal-00979685〉

Partager

Métriques

Consultations de la notice

243